5.4 Supply voltage
The transducer must be operated with a separated extra‐low voltage (nominal
(rated) supply voltage 18 30 V
flanges within a test bench at the same time. Should the device be operated
on a DC voltage network
excess voltages.
The information in this section relates to the self‐contained operation of the
T40B, without HBM system solutions.
The supply voltage is electrically isolated from signal outputs and shunt signal
inputs. Connect a separated extra‐low voltage of 18 V 30 V to pin 3 (+) and
pin 2 ( ) of connectors 1 or 3. We recommend that you use HBM cable KAB
8/00-2/2/2 and the appropriate sockets (see Accessories). The cable can be
up to 50 m long for voltages 24 V, otherwise it can be up to 20 m long.
If the permissible cable length is exceeded, you can supply the voltage in par
allel over two connection cables (connectors 1 and 3). This enables you to
double the permissible length. Alternatively, install a power supply on site.
Important
The instant you switch on, a current of up to 4 A may flow and this may switch
off power packs with electronic current limiters.
1)
Distribution system for electrical energy with greater physical expansion (over several test benches, for
example) that may possibly also supply consumers with high nominal (rated) currents.
HBM
). You can supply one or more torque
DC
1)
, additional precautions must be taken to discharge
T40B