Herunterladen Inhalt Inhalt Diese Seite drucken

Mounting The Rotor - HBM T40B Montageanleitung

Vorschau ausblenden Andere Handbücher für T40B:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

16

4.5 Mounting the rotor

Tip
Usually the rotor type plate is no longer visible after installation. This is why
we include with the rotor additional stickers with the important characteristics,
which you can attach to the stator or any other relevant test‐bench compon­
ents. You can then refer to them whenever there is anything you wish to know,
such as the shunt signal. To explicitly assign the data, the identification num­
ber and the size are engraved on the rotor flange, where they can be seen
from outside.
1. Prior to installation, clean the plane faces of the transducer flange and the
counter flange.
For safe torque transfer, the faces must be clean and free from grease.
Use a piece of cloth or paper soaked in solvent. When cleaning, make sure
that you do not damage the transmitter winding or the rotational speed
measuring system.
Transmitter winding
Without rotational speed
measuring system
Fig 4.1:
Bolted rotor connection
HBM
Rotational speed measuring system
Flange plane faces
T40B
With rotational speed
measuring system
A3452-4.0 en/de

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis