8
Fault – Cause – Remedy
Fault / message
Not possible to switch on the instrument
The instrument cannot be charged.
Notice: "Instrument failure" or "Sensor has
bad contact"
Instrument displays potentially individually
incomprehensible characters
Instrument turns on continually
The instrument is defective
Flow too low!
No valid span calibration
Dräger X-am 7000
Cause
Power pack is fully discharged.
Temperature too high or too low
Power pack is not connected correctly or bad
contact in the power-pack connector.
Wrong language selected
Electronic fault (Watchdog)
An internal fault has been detected.
#2: Instrument fault
#129: Electronic fault
#130: Fault when switching on the flow sensor
Hose is blocked
#9 Pump faulty
#107: Calibration was executed without test gas Repeat calibration with test gas.
Fault – Cause – Remedy
Remedy
Charge the power pack.
Note: In this case, the instrument display
initially remains blank.
Observe temperature range.
Reconnect the power pack and secure it with
the screws.
Select the correct language with the program
Dräger CC Vision. Alternatively, navigate
"blindly" through the menu structure. Language
is the bottom option (see page 64).
Menu: (3rd option) Settings\ (2nd option)
Instrument \ (last item) Language.
Switch the instrument off then on again, or
remove and reconnect the power pack. If the
fault is still displayed, contact DrägerService.
Switch the instrument off then on again, or
remove and reconnect the power pack. If the
fault is still displayed, contact DrägerService.
Inspect the hose for kinks and blockages.
If necessary, remove the pump adapter and
inspect the membrane in the sensor grille for
dirt and drops of water.
Ensure that the hose is not kinked or
submerged during use.
Replace the pump.
73