Herunterladen Inhalt Inhalt Diese Seite drucken

Canadian Installation; Common Instrument Power Cord Plugs - Agilent Technologies ALS Handbuch

Inhaltsverzeichnis

Werbung

To protect users, the metal instrument panels and cabinet are grounded
through the three-conductor power line cord in accordance with International
Electrotechnical Commission (IEC) requirements.
The three-conductor power line cord, when plugged into a properly grounded
receptacle, grounds the instrument and minimizes shock hazard. A properly
grounded receptacle is one that is connected to a suitable earth ground. Be
sure to verify proper receptacle grounding. The GC requires an isolated
ground.
Connect the GC to a dedicated circuit.

Canadian installation

When installing a GC in Canada, make sure your GC's power supply circuit
meets the following additional requirements:
• The circuit breaker for the branch circuit, which is dedicated to the
• The service box branch circuit is marked as a "Dedicated Circuit."

Common instrument power cord plugs

Table 33
Table 33
Power cord terminations
Country
Australia
GC, GC/MS, and ALS Site Preparation Guide
instrument, is rated for continuous operation.
below shows common Agilent power cord plugs.
Voltage
Amps
240
16
Cable
length (m)
Wall termination
2.5
AS 3112
6850 Series GC Site Preparation
Plug
Termination
75

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

MsGc

Inhaltsverzeichnis