7890 Series GC Site Preparation
Table 12
Delivery pressures for inlets required at the GC/MS, in kPa (psig)
Inlet type
Split/Splitless
150 psi
Carrier (max)
1,172 (170)
Carrier (min)
(20 psi) above maximum pressure used in method. (If using constant flow control in the inlet, the maximum
column pressure occurs at the final oven temperature.)
1 Japan only: 1013 (147)
Table 13
inlets and detectors, measured at the bulkhead fittings on the back of the
instrument.
Table 13
Carrier (max)
Carrier (min)
1 Japan only: 1013 (147)
Table 14
Maximum delivery pressures for detectors, at the GC/MS, in kPa (psig)
Detector type
FID
Hydrogen
240–690 (35–100)
Air
380–690 (55–100)
Makeup
380–690 (55–100)
Reference
36
Split/Splitless
100 psi
1
827 (120)
and
Table 14
list minimum and maximum delivery pressures for
Delivery pressures for inlets required at the GC/MS, in kPa (psig)
Inlet type
Split/Splitless
150 psi
1,172 (170)
(20 psi) above maximum pressure used in method. (If using constant flow
control in the inlet, the maximum column pressure occurs at the final oven
temperature.)
NPD
240–690 (35–100)
380–690 (55–100)
380–690 (55–100)
Multimode
On-column
100 psi
1,172 (170)
827 (120)
Split/Splitless
100 psi
1
827 (120)
TCD
ECD
380–690 (55–100)
380–690 (55–100)
380–690 (55–100)
GC, GC/MS, and ALS Site Preparation Guide
Purged
PTV
packed
827 (120)
827 (120)
Multimode
100 psi
827 (120)
FPD
310–690 (45–100)
690–827 (100–120)
380–690 (55–100)