12.0 INSTRUCTIONS WITH REGARD TO EMC

EN

12.1 GUIDELINES AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY

Der NEBUFIRST
®
The customer or user of the NEBUFIRST
Electromagnetic
immunity tests
Conducted disturbances
as per IEC 61000-4-6
Radiated HF disturbances
as per IEC 61000-4-3
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not be applicable in all cases. The propagation of electromagnetic values is influenced by absorptions
and reflections of buildings, objects and people.
a
The field strength of stationary transmitters such as base stations of wireless phones and land mobile radio units,
amateur radio stations, AM and FM radio and television transmitters, cannot be predetermined exactly in theory.
To establish the electromagnetic environment in regard to stationary transmitters, a study of the location should be
considered. If the field strength measured at the location where the NEBUFIRST
above accordance levels, the NEBUFIRST
as intended. If unusual performance characteristics are observed, additional measures may be required, such as for
instance a change of orientation or a relocation of the NEBUFIRST
b
Above the frequency range of 150 kHz to 80 MHz, the field strength should be below 3 V/m.
INHALER is intended for operation in the electromagnetic environment specified below.
INHALER should ensure that the device is used in such an environment.
®
IEC 60601- test level
3 V
eff
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Accordance level
10 V
10 V/m
INHALER should be observed in order to verify that the device functions
®
INHALER.
®
42
Electromagnetic environment -
guidelines
Portable and mobile radio devices should
not be used at a distance closer to the
NEBUFIRST
INHALER device, including
®
the cables, than the recommended safety
distance as calculated with the equation
that applies to the transmission frequen-
cy.
Recommended safety distance:
d = 0.35 √ P
d = 0.35 √ P
80 MHz to 800 MHz
d = 0.70 √ P
800 MHz to 2.5 GHz
where P is the nominal power of the
transmitter in watt (W) according to the
specifications of the manufacturer of the
transmitter and d is the recommended
safety distance in metres (m).
The field strength of stationary radio
transmitters should be lower than the
accordance level (b) for all frequencies
as determined by an on-site investigation
(a).
Interference is possible in the vicinity
of devices marked with the following
pictogram.
INHALER is used exceeds the
®