6.2.4.

Listing (reconsider)

All measured data appears in a numeric list. The
date beside the time gives more exactly values
to see what happened on time units smaller 1 s.
The value specifies the time in seconds after
midnight (0:00 h). The amount of data depends
on the frequency that readings were taken (set-
tings at 12.1. time interval between readings).
As the amount of data increases, so does the
amount of storage space required to save it. In
order to save room, all .i12 data files are stored
by a binary code.
Note:
Data files may be converted into text files (.txt file: in the menu bar under output
ASCII file save as .txt file), and can be easily opened with EXCEL. EXCEL formats
the columns accordingly, if you specify that the headers should be ignored.
6.2.5.

Trend output (reconsider)

The graph's curve depicts the temperature
change over time within the specified tempera-
ture range. Additionally, other information ap-
pears in this window; such as recorded time (x-
axis) and temperature in degrees (y-axis) as
well as the time and temperature at the vertical
cursor line which can be dragged with the
mouse.
Selecting the Trend output initially causes all
the saved data to be displayed.
If the data exceeds an amount that can be rep-
resented reasonably, you may "Zoom" in on a
partial segment using the mouse (such as the
segment represented in the example). Under
"Total" you can return to the representation of
the entire curve.
Note:
Data files may be converted into text files (.txt file: in the menu bar under output
ASCII file save as .txt file), and can be easily opened with EXCEL. EXCEL formats
the columns accordingly, if you specify that the headers should be ignored.
Note:
There is a description of the individual icons in the program's help menu. Click on the
F1 key after InfraWin has been loaded or click on the ? in the menu bar.
7.
Interface commands analog measuring mode
The data exchange occurs in ASCII format with the following transmission parameters:
The data format is:
The device responds with the entry of a command with: output (e.g. the measuring value) + CR (Carriage
Return, ASCII 13), with pure entry commands with "ok" + CR.
Every command starts with the 2-digit device address AA (e.g. "00").
This is followed by 2 small command letters (e.g. "ms" for measuring value)
This is followed, if necessary for that command, by the ASCII parameter "X". If this parameter "X" is omitted,
then the device resets with the current parameter.
Example: Entry: "01ms" + <CR>
The current measuring temperature of the device is given back with the address 01
Answer: "07568" + <CR> means 756.8 °
8 data bits, 1 stop bit, equal parity (8,1,e)
[aMes]