Device setup

4 Device setup
We recommend performing the configuration steps in an office before mounting the device, especially
if you are not familiar with the LLB-60-D.
The LLB-60-D supports two types of operation modes:
Controlled mode
Automatic mode (for use in stand-alone operation)
The first decision to be taken is the type of operation mode that will be used to transmit the distance
measurement data. While the controlled mode provides maximum flexibility and accuracy, it is often
not suitable for integration into existing PLCs or analog environments. In such cases the automatic
mode might be preferred.
Influence
Accuracy
Maximum measurement accuracy
Flexibility
Access to full command set
Integration
Requires protocol implementation
Connection of up to 10 LLB-60-D to a

Connection

single RS-422 line.
The following two sections describe the configuration of the LLB-60-D for the controlled and automatic
modes.
4.1 Connection
To be able to configure the LLB-60-D, it must be powered and connected to a PC. Figure 2 shows the
necessary connections. On the PC, any terminal program can be used to communicate with the
module. A configuration utility is also available on the web page www.tr-electronic.de.
Figure 2: Connection for LLB-60-D configuration
TR-Electronic GmbH 2007, All Rights Reserved
Page 66 of 96
controlled mode
TR - ELE - BA - DGB - 0015 - 02
automatic mode
(with analog output and/or digital outputs)
Accuracy depends on signal scaling
(see 10.3.5 Set/Get distance range (sNv)
on page 86)
Limited
Wiring of AO and/or DO signals
Point-to-point connection
Printed in the Federal Republic of Germany
07/22/2009