Tintometer Lovibond MD 6x0 Gebrauchsanleitung Seite 18

Werbung

Mode
Description /
Number
Display
40
Calibration
45
User
Calibration
EN
46
Clear
Calibration
50
Profi Mode
55
One Time
Zero
60
Method list
61
M list all on
62
M list all off
64
User
concentration
18
4 Operation
How to Execute function
• Select method from list for calibration and press Enter
key to confirm.
• Refer to chapter 4.5. for detailed procedures.
• Press Shift + 2 keys to reset method blank calibration of
Method 191 to factory Calibration.
• Select mode when test results are displayed.
• Press Arrow keys to increase or decrease test results to
match the corresponding Standard values.
• Press Enter key to save the results.
• Refer to chapter 4.5. for more information.
• Select mode before zero measurement of required
method.
• Press Shift + 0 keys to retain user calibration.
• Press Shift + 1 keys to delete user calibration.
• Press Shift + 0 keys to switch off profi mode.
• Press Shift + 1 keys to switch on profi mode.
• Press Enter key  to confirm.
Refer to chapter 4.5. for more details.
• Press Shift + 0 keys to switch off OTZ.
• Press Shift + 1 keys to switch on OTZ.
• Press Enter key to confirm.
Refer to chapter 4.5. for more details.
This setting allows user to create a user defined list of
favorite methods. The program structure requires that the
list must have at least one active method. For this reason,
it is necessary to activate all methods and then deactivate
the ones which are not required.
• Press Enter key to display favourite methods.
• Press Arrow keys to select desired method.
• Press F2 key to activate or inactivate specific method
and press Enter key to confirm.
• Methods with [∗] behind method number will be displayed
only.
• Press Shift + 1 keys to switch all user methods on.
• Press Shift + 0 keys to keep valid methods list.
• Press Shift + 1 keys to display only one method on the
list.
• Press Shift + 0 keys to keep valid methods list.
• Create a user method by entering concentration values.
• Refer to chapter 4.5. for detailed instructions.
EN   Photometer Series