Function check

6.1 Intervals

6.2 Performing a function check

Result
Carry out a function check at regular intervals:
After each hygienic preparation
After each repair
At least every 6 months
1. Check whether all the modules are set up and connected
correctly in accordance with the instructions for use.
2. Inspect the housing for cracks, damage, and heavy soiling.
3. Connect the module up to the therapy device
Connecting up the prismaPSG module", page
4. When using with prismaCONNECT: Check whether the
symbol is shown on the therapy device display.
When using with prisma HUB: Check whether the prismaPSG
LED lights up on thecontrol panel membrane.
5. Check whether a value corresponding to the information on
the therapy device display is shown on the pressure channel
being fed-in.
6. If the module does not work properly or shows signs of
damage: Contact your authorized dealer.
The function check is complete.
6 Function check
(see "4.1
30).