4 Performing a functional test
4.3 Executing a "Pan symmetry" test exposure Converting instructions for the "ORTHOPHOS XG 5 to XG 3D upgrade kit" ORTHOPHOS
14
A
Adjustment OK
B
C
Adjustment not OK
2. Examine the exposure.
3. If there is something wrong with the exposure, inform the operator.
NOTICE! The conversion may only be performed on a unit that is
intact.
Sirona Dental Systems GmbH
● The shadow of the center needle,
the needle image, and the auxiliary
lines must be coincident and located
one behind the other. A tolerance
(offset of needle from the central
auxiliary line) of ± 0.75 mm is
admissible.
● Distance A1 must be 88.6 ± 1 mm.
● Distances A2 must be identical,
each being 44.3 ± 0.5 mm.
● A surrounding white border must be
present.
D3352.092.08.05.02 07.2012
63 52 970 D3352