Information on Converting the Device
11.1
Changing the concentration range to 6...10%
Where concentrations are greater than 5%, the maximum water flow rate must be reduced.
This entails replacing the flow regulator.
Flow value
360 l/h
840 l/h
Proceed as follows to change the concentration range:
 Pull out the mains plug and interrupt the water infeed (close the angle valve).
 Pull out plug (Item 1) and loosen nut (Item 2)
 Loosen swivel nuts (Item 3), remove solenoid valve (Item 4)
 Remove old flow regulator (Item 5) with a suitable tool
 Insert the new flow regulator (Item 6). You must make sure that the new flow regulator is
positioned correctly (see Fig. 11.2 for the installation position).
 Reinstall the solenoid valve in reverse order. Ensure that the flat seal is correctly seated.
 Reconnect the electricity and water supply.
 Check that the device is functioning properly and make sure that it is not leaking.
417102242_DG_3
Concentration range
max. 10 % (0.25-10%)
max. 5 % (0.25-5%)
Item
1
2
3
4
5
6
Fig. 11.1
Fig. 11.2

Flow regulator installation position

- 91 -
Comment
Separate accessory
Standard
Description
Plug
Nut
Swivel nut
Solenoid valve
Old flow regulator
New flow regulator

Corner valve retrofit

Rev.: 02-05.2014