The first one is set by the doctor when the device will be used by the patient for homecare use; the second mode instead allows full
access to all the functions of the device when used directly by the doctor.
Select the desired mode, press the OK icon, automatically the mode will be set and the device will return to the service menu. The service
menu will show a different configuration according to which mode has been selected.
For more information concerning device functioning in patient mode please view paragraph 2.7.2
Change date/time
Select the item by touching the display.
When setting the date and time, the cursor _ indicates the data item which is being modified. Use the numbers shown to modify the data
item of interest, move on to the next data item by touching OK. Touch OK so that the new settings will take effect and to return to the
service menu. To return to the service menu without modifying the item data touch
LCD Settings
This item allows to:
•
Brightness and contrast setting of the display
With two scales which go from 0 to 31 it is possible to set the parameters of the display and view in real time its effect. Once the
best combination of brightness and contrast is obtained touch OK on the bottom right side of the display
•
Calibrate the touch screen function
This function is used to check the proper response of the touch screen; the device initially request a confirmation, touching OK
icon the calibration begin
There are four phases as follows:
•
Touch three times the top left-hand point of the display
•
Touch three times the top right-hand point
•
Touch three times the bottom right hand point
•
Touch three times the bottom right hand point
The reference area is the one inside the flashing outline.
This procedure will then be set up in function of the dimensions of the display.
This procedure must be carried out using the tip of a touch-screen "pointer" held vertically at 90 degrees to the display.
if the calibration is made correctly then the device shows:
Otherwise the user is requested to repeat the calibration procedure.
The procedure cannot be cancelled once started, so carry out the procedure correctly to return to the service menu.
Bluetooth Settings
Once in the menu it is possible to select the activation mode of the Bluetooth function. The "Activation" item allows to select from the
following options; "On request" and "always on"; in the first case the function is activated only when requested(for example to print a
test),otherwise it remains off thus allowing to save energy; by selecting the option "Always on" this function is always activate and ready
for use(for example to transfer data to a mobile phone.)
Access this menu to search for active Bluetooth devices, touch the option "Search Device"; Spirodoc will start to search for Bluetooth
devices in the area; once one or more devices are found the display will list these devices with their respective names. By touching the
device of interest it will be memorized as a printer, a phone, or as a PC – On line; select an option.
In the "Bluetooth setting" menu any previously memorized devices can be viewed in the "printer" list, "telephone" list or the "PC – On
line" list. Next to the device name a corresponding icon (telephone, printer or PC) will appear. Any device from these lists can be set as
the default device (the device that Spirodoc will automatically connect to via Bluetooth) by entering in the lists, touching the display and
selecting the device. A listed device can be eliminated from the list. (in this specific case the user will confirm the deletion with the OK
icon.)
So as not to make any modifications touch the
Turn ON mode
This function enables the device to turn on automatically at a predetermined time. The device will automatically turn on and commence
a sleep oximetry test (this test is also capable of monitoring the patient all day long, and includes as a step counter and a triaxial
accelerometer to measure the VMU)
The device will turn off automatically at a predetermined time.
SPIRODOC
Calibration is OK
icon in the bottom left side
cod. 980156
Rev 4.2.1
.
EN 25/48
spirodoc
User manual