Englisch
2.9.1.1.22.2 command: tbpos
type
syntax
parameter
description
range
dimension unit
default value
request
response
error
example
back to command list
2.9.1.1.22.3 command: tblo
type
syntax
parameter
description
range
dimension unit
default value
request
81
Geschäftsführer: Dr. Bernt Götz • Telefon 03641/66880 • Fax 03641/668866 • www.piezojena.com
content
tbpos,ptr[,slew rate,destination,duration]
ptr
table pointer
slew rate
slew rate
destination end voltage
duration
duration, total time of the table section (after this period ends this section of
the table even if the slew rate is not yet expired).
It is a full table section ( slew rate,destination,duration ) is written table pointer ptr to the.
If the command is called only with the parameter ptr, it returns the contents of the table
pointer. When a fault occurs in the table section is not changed.
ptr
0 ... 99
slew rate
0.005 ... 0.000000003
destination 0 ... 100
duration
0.1 ... 100
ptr
-
slew rate
V/µs
destination %
duration
s
slew rate
0.005
destination 0
duration
0.1
write:
tbpos,ptr,slew rate,destination,duration <CRLF>
read:
tbpos,ptr <CRLF>
read (tbpos,ptr <CRLF> ):
tbpos,ptr,slew rate,destination,duration <CRLF> - if no error
or
nok <CRLF> - if error
write (tbpos,ptr,slew rate,destination,duration <CRLF> ):
ok <CRLF> - if no error
or
nok <CRLF> - if error
error-bit 28, error-bit 29, error-bit 31
message: tbpos,0,0.0003,50,50 <CRLF>
response: ok <CRLF> - if no error
or
nok <CRLF> - if error
content
tblo[,lowptr]
lowptr
It defines the lower limit for the output pointer. Lowptr exceeds the currently set output
current pointer and will be made lowptr .
lowptr
0 ... 99
-
lowptr
0
write:
tblo,lowptr <CRLF>
read:
tblo <CRLF>