Herunterladen Inhalt Inhalt Diese Seite drucken

Appendix; Technical Data - Lowenstein Medical prisma CHECK Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

10 Appendix

10.1 Technical data

Specification
prisma CHECK
Product class according to 93/42/EEC
IIa
Dimensions W x H x D in cm
4.3 x 12.8 x 8
Weight in kg
0.185
Temperature range
Operation
+5°C to +40°C
Storage
-25°C to +70°C
Permissible humidity during operation and
Rel. humidity 10% to 90%, non-condensing
storage
700 hPa to 1060 hPa, corresponds to a height of
Air pressure range
approx. 3000 m above sea level
Power supply
12 V DC
Electrical output
1.2 VA*
Insulation of remote alarm interface
1.0 kV
Switching capacity of remote alarm interface
60V DC/2A; 42V AC/2A
0% to 100%
SpO
display
2
Pulse display
0 bpm to 300 bpm
Update period for SpO
and pulse
0.5 s
2
Applied part
SpO
sensor
2
Classification as per DIN EN 60601-11:
Protection class II
Type of protection against elec. shock
Degree of protection against elec. shock
Type BF
Protection against harmful ingress of water and
IP21
solid particles
Classification as per DIN EN 60601-1:
Continuous operation
Operating mode
Electrical medical devices may only be installed and
commissioned in a defined electromagnetic
environment with regard to emission and radio
interference immunity. More information, including
Electromagnetic compatibility (EMC) as per
EN 60601-1-2
test parameters and limit values, can be obtained from
Radio interference suppression
the manufacturer if required.
Radio interference immunity
EN55011 B
IEC 61000-4 Parts 2 to 6, Part 11, Part 8
IEC 61000-3 Parts 2 and 3
prisma CHECK