Possible problem
Dark or over-
illuminated image.
The colour tone of the
endoscopic image is
unusual.
Picture is frozen.
6.6. Remote switches/programmable buttons
Possible problem
The remote switches
are not working or not
working properly.
20
Possible cause
aBox 2 image settings
incorrect.
aScope Gastro is
defective.
aScope Gastro is
defective.
aScope Gastro is
defective.
aBox 2 is defective.
Possible cause
aScope Gastro connector
is not properly connected
to the aBox 2.
Remote switch
configuration changed.
Wrong remote switch
operated.
aScope Gastro is
defective.
aBox 2 is defective.
Recommended action
See aBox 2 IFU.
Withdraw the aScope Gastro
and connect a new endoscope.
Withdraw the aScope Gastro
and connect a new endoscope.
Withdraw the aScope Gastro
and connect a new endoscope.
Contact your Ambu
representative.
Recommended action
Connect the aScope Gastro
connector properly to the
aBox 2.
Return to standard
configuration of the remote
switches or change the settings.
Operate the correct remote
switch.
Withdraw the aScope Gastro
and connect a new endoscope.
Contact your Ambu
representative.