Error/Message
Device out of tem-
perature range.
↑ ↑ ↑
(over measuring
range)
Insufficient volume of
breath. [Only for
breath samples using
the mouthpiece]
Volume of breath too
large. [Only for breath
samples using the
mouthpiece]
Sample disruption
Error number
Dräger Alcotest
Cause
Device is too warm or
too cold.
Measuring range
exceeded. Residual
mouth alcohol can
distort the measured
values.
Volume of breath
sample insufficient.
Measuring range
exceeded.
Irregular exhalation,
abrupt cessation or
inhalation.
Device error.
®
7000 med
Troubleshooting
Remedy
Allow the device to
cool down or warm
up.
A wait time of at least
15 minutes after alco-
hol was last con-
sumed must have
elapsed. Press the
button to repeat.
Repeat the breath
sample after pressing
the
button.
If necessary, carry out
a passive breath sam-
ple.
Repeat the breath
sample after pressing
the
button.
Repeat the breath
sample after pressing
the
button.
If necessary, carry out
a passive breath sam-
ple.
Turn the device off
and then back on. If
the error persists,
contact DrägerService
or your dealer.
|