3.2. Remote control commands
120
Assembly.Outputs.SetLines
Assembly.Outputs.SetLines.L0
up to .L 3
Assembly.Outputs.ResetLines
Setting the I/O output lines.
.AutoEOD: The automatic output of the EOD (End of Determination) at the
end of the determination can be switched off. Thus, for example,
in conjunction with a Titrino several determinations can be per-
formed in the same beaker. Before AutoEOD is switched on, line 3
must be set to "OFF".
.SetLines: With $G, all lines are set.
.SetLines.LX: Set the line LX. "active" means setting of a static signal, "inac-
tive" means resetting of the signal, "pulse" means output of a
pulse of app. 150 ms, "OFF" means the line is not operated, see
also page 162.
Warnings:
• L2 is the EOD line. If you have "AutoEOD" set to "ON", an active
line 2 is set to "inactive" by the EOD pulse.
• L3 is the line of the activate pulse. An active line 3 is set to "in-
active" by the activate pulse.
.ResetLines: Lines are set to the inactive status (= high).
3.2.2.88. Setup.Keycode
ON means the key code of a key pressed on the Titrino is outputted. The key
code comprises 2 ASCII characters; table of the keys with their code, see
page 139. A keystroke of key 11 is sent as follows:
#11
The beginning of the message is marked by a space (ASCII 32).
3.2.2.89. Setup.Tree.Short
Setup.Tree.ChangedOnly
Definition of the type of answer to $Q.
.Short:
With "ON", each path is sent with only the necessary amount of
characters in order to be unequivocal (printed in bold in this manual). A com-
bination of .Short and .ChangedOnly is not possible.
.ChangedOnly:
Sends only the changed values, i.e. values which have
been edited. All paths are sent absolute, i.e. from the root.
3.2.2.90. Setup.Trace
The Titrino automatically reports when a value has been confirmed with <en-
ter> at the Titrino. Message, e.g.:
&SmplData.OFFSilo.Id1"Trace"
The beginning of the message is marked by a space (ASCII 32).
3.2.2.91. Setup.Lock.Keyboard
Setup.Lock.Config
Setup.Lock.Parameter
Setup.Lock.SmplData
active, inactive, pulse, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF