Specifications
Model:
Measurement Method
Display
Measuring range
Accuracy
Automatic Power Off
Battery Type
Battery life
Storage condition
Operation condition
Dimensions (approx.)
Weight (approx.)
For purposes of improvement, specifications are subject to change without notice
Troubleshooting
Error/Symbol
Shows abnormal
result
ENGLISH
DXJ-210
Measurement: auscultation methods, methods of oscillographic
liquid crystal digital display
Pressure: 0~300 mmHg
Pressure: within ± 3mmHg
About 5 minutes after measurement is complete
2 AA Batteries
About 2 months with 3-min. usage per day
(14~131) °F, (10-95) %RH
(41~104) °F, (30~85)%RH
Monitor: 360x96x66 mm
Cuff: fits wrists 22-33 cm
1100g (including batteries)
What does It Mean?
Battery low
Cuff not tightened properly or its posi-
tion is incorrect
Body is not steady
Speaking, frightened nervous or are
excited during measurement
Wrong position
Some interference in inflation or
wrong operation during measuring
How To Fix It
Change the battery
Read pages 10 (Important Reminders
Stay calm and remain steady
Do not speak and take deep breath
2~3times to relax yourself
Adjust position; refer to "Usage".
Refer to the inflation step in "Taking a
Blood Pressure Measurement'.