Trouble Shooting
E3
E3
Slope or asymmetry
not within the
admissible ranges
(Slope:
-50 ... -62 mV/pH;
Asymmetry:
-30 ... +30 mV)
No message
Electrode voltage
not stable
No message
Wrong measured
values
Instrument doesn't
react upon key
operation
LoBat
to
to
pH electrode:
pH electrode too old
Diaphragm contaminated
Membrane contaminated
Humidity in the plug
Old electrolyte solution
Instrument:
Wrong calibration
procedure
Wrong solution temperature
(without temp. sensor only)
Humidity in the socket
Buffer solution:
Improper buffer solutions
Spoiled buffer solutions
pH electrode:
Diaphragm contaminated
Membrane contaminated
Sample solution:
ph-value not stable
Temperature not stable
pH electrode +measuring
solution:
Too low conductivity
High temperature
Organic liquids
pH electrode:
Improper pH electrode
Temperature difference
between calibration and
measuring solution too high
Interference voltages
Improper measuring
procedure
Undefined operation
condition of the processor e.g.
after non-admissible EMC
admission
Rechargeable batteries nearly
discharged.
Timeout of the serial interface Check connected instrument
pH electrode:
Replace pH electrode
Clean diaphragm
Clean membrane
Dry plug
Exchange electrolyte solution
Instrument:
Select right procedure
Set right temperature
Dry socket
Buffer solution:
Select solutions corresp. to the
calibration procedure
Use solutions once only
pH electrode:
Clean diaphragm
Clean membrane
Sample solution:
Possibly measure without air
contact
Possibly temper
Use suitable pH electrode
Use suitable pH electrode
Use suitable pH electrode
pH electrode:
Select suitable pH electrode
Temper buffer or measuring
solutions
Do not connect grounded PC
or printer if measuring solution
is grounded
Attend to special procedures
Processor reset:
Press key RCL and switch on
the instrument
Charge.