Herunterladen Inhalt Inhalt Diese Seite drucken

Manual Cleaning With Wipe Disinfection - B. Braun Aesculap Acculan 4 GA344 Gebrauchsanweisung/Technische Beschreibung

Kleine bohrmaschine
Vorschau ausblenden Andere Handbücher für Aesculap Acculan 4 GA344:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Aesculap
Acculan 4 small drill GA344
7.6

Manual cleaning with wipe disinfection

Phase
Step
I
Pre-cleaning
II
Cleaning with enzyme
solution
III
Intermediate rinse
IV
Drying
V
Wipe disinfection
VI
Final rinse
VII
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
RT:
Room temperature
*
validated with enzyme cleaner "Cidezyme Johnson & Johnson"
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
Phase I
Clean the product under running water, using a suitable cleaning brush
until all visible residues have been removed from the surfaces.
Brush non-visible surfaces (such as cannulation) with a cleaning brush
TA011944.
Move flexible components (such as sleeves) during cleaning.
Phase II
Follow the operating instructions of the enzyme cleaner with regard to
correct concentration, dilution, temperature and water quality.
Spray products with a pH neutral enzyme solution, let soak in for at
least 2 minutes and then wipe off.
Phase III
Rinse product under running tap water for at least 5 minutes.
Move flexible components (such as sleeves) during cleaning.
Follow the operating instructions of the enzyme cleaner with regard to
correct concentration, dilution, temperature and water quality.
Remove contamination with a lint-free cloth or soft brush moistened
with enzyme cleaner.
Rinse flexible components (such as sleeves) and cannulations for 20
seconds with the water pistol (cold water, at least 2.5 bar).
After manual cleaning, check visible surfaces and areas of flexible
components for residues.
If necessary, repeat the cleaning process (phase 1 to 3).
10
Acculan 4
®
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold) ≥2
-
RT (cold) ≥2
0.8
≥5
RT
-
RT
-
-
-
>1
-
RT (cold) 0.5
-
RT
-
-
Water quality
Chemicals
D-W
Until visually clean
D–W
pH-neutral*
D-W
-
-
-
Meliseptol HBV wipes 50 % Propan-1-ol
FD–W
-
-
-
Phase IV
Dry the product in the drying phase with suitable equipment (such as
lint-free cloths, pressurized air).
Phase V
Wipe all surfaces of the product with a single-use disinfecting wipe.
Phase VI
Rinse disinfected surfaces after the prescribed reaction time for at
least 1 minute under running demineralized water.
Drain any remaining water fully.
Stage VII
Dry the product in the drying phase with suitable equipment (such as
lint-free cloths, pressurized air).