Site Requirements for the Agilent 2100 Bioanalyzer Instrument
Physical Specifications of the 2100 Bioanalyzer Instrument
Physical Specifications of the 2100 Bioanalyzer
Instrument
Table 1
Bioanalyzer instrument.
Table 1
Physical specifications
Type
Weight
Dimensions
Line voltage
Line frequency
Power consumption
Ambient operating temperature
Storage temperature
Operating humidity
Operating altitude
Safety standards: IEC, EN,CSA,
UL
ISM Classification
Sound pressure
on page 115 lists the physical specifications of the Agilent 2100
Specification
10 kg (22 lbs)
162 — 412 — 290 mm
(6.4 — 16.2 — 11.4 in)
100-240 VAC ±10%
50-60 Hz ±5%
30 W / 60 VA
15-27 °C (59-80 °F)
-40-70 °C (-40-158 °F)
<70%, at 15-27 °C (59-81 °F)
Up to 2000 m (6500 ft)
Overvoltage Category II, Pollution
Degree 2
ISM Group I Class B:
CAN ICES-001(B)/NMB-001(B)
<70 dB (A)
Comment
Width — depth — height
Wide-ranging capability
The assays set-up to be analyzed on a chip
temperature of 30 °C. If the ambient
temperature is above 30 °C the assays are
analyzed using that temperature.
Non-condensing
For indoor use only.
According to CISPR 11
According to DIN-EN 7779
115