5. Commissioning, operation
5.2.3 Connection terminals and wire preparation of the switch contacts
EN
4-pin terminal: Connection of the transmitter
12-pin TTI plug-in terminal: Connection of the switch contacts
CAUTION!
Damage to the instrument through incorrect connection
Connecting the switch contacts or the transmitter to the wrong cable termi-
nals can lead to irreversible damage to the instrument.
The connection terminals are suitable for the following types of single conductor or cross-
sections:
Connection terminals and wire preparation
Types of single
conductors
Wire length
Maximum length of
the bare wire end
Number of wires /
Cross-sections
Recommended
tightening torque
16
Ensure that the pin assignment is correct.
Connection terminals of
the socket terminal strip
for the switch contacts
Solid wire core
Flexible stranded wire
Flexible stranded wire with end
splice
≥ 90 mm
Max. 5.5 mm
1 x 0.5 mm² to 1 x 2.5 mm²
0.5 Nm
Connection terminals of
the protective conductor
Solid wire core
Flexible stranded wire with end
splice
≥ 120 mm
Max. 8 mm
1.2 Nm
WIKA operating instructions, model GDM-RC-100-T