Herunterladen Inhalt Inhalt Diese Seite drucken

Commissioning, Operation; Mechanical Mounting - WIKA GDM-RC-100-T Betriebsanleitung

Hybrid-gasdichtewächter mit referenzkammer
Vorschau ausblenden Andere Handbücher für GDM-RC-100-T:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

5. Commissioning, operation

5. Commissioning, operation

5.1 Mechanical mounting

EN
CAUTION!
Physical injuries and damage to property and the environment
through faulty instrument
Prior to commissioning, the instrument must be subjected to a visual
inspection.
5.1.1 Requirements for the installation point
For outdoor applications, the selected installation location has to be suitable for the
specified ingress protection, so that the instrument is not exposed to impermissible
weather conditions.
The sealing faces at the instrument and at the measuring location have to be undam-
aged and clean.
The measuring instruments must be mounted in the common mounting position per
EN 837-1, with a max. permissible incline of 5° on all sides.
The measuring location should preferably be positioned directly at the
gas compartment. A measurement at the end of measuring lines prevents
optimal results through unwanted temperature differences to the main tank.
5.1.2 Installation
With transport or storage, it can occur that gas density monitors warm up or cool down
and this results in pointer movements. These pointer movements are caused by the
compensation system. To make sure that the instruments have adapted sufficiently
to ambient temperature, at least 2 hours at 20 °C [68 °F] must be allowed for adapta-
tion to the temperature. Then, in the depressurised state, the pointer will sit within the
tolerance bar.
Corresponding to the general technical rules for pressure gauges (e.g. EN 837-2
"Selection and installation recommendations for pressure gauges") when screwing in
the instrument, the force required to do this must not be applied through the case, but
only through the spanner flats provided for this purpose and using a suitable tool.
When screwing in, do not cross the threads.
For parallel threads, use flat gaskets, lens-type sealing rings or WIKA profile sealings at the
sealing face . With tapered threads (e.g. NPT threads), sealing is made in the threads ,
using a suitable sealing material (EN 837-2).
The tightening torque depends on the sealing used. In order to orientate the measuring
instrument so that it can be read as well as possible, a connection with LH-RH adjusting
nut or union nut should be used. When a blow-out device is fitted to an instrument, it must
be protected against being blocked by debris and dirt.
14
Only use the instrument if it is in perfect condition with respect to safety.
WIKA operating instructions, model GDM-RC-100-T

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis