5. Commissioning, operation
5.4.2.2 Writing new parameters
Take note of the new communication parameters before writing them, as the
parameters will be required again for any new access to the transmitter.
EN
Write the new values in the right fields (below the button "Write on instru-
ment").
Designation
TAG number (name of instrument)
Address
Baud rate
Parity
Gas constituents (freely selectable
from SF₆, N₂, CF₄, O₂, CO₂,
3M™ Novec™ 4710, He and Ar)
By pressing the button "Write on instrument" the data in the fields is transmit-
ted to the instrument register. To finish the writing operation, interrupt the
voltage supply of the transmitter after the transmission before restoring it.
Afterwards, during the reading operation, the entered data becomes
visible on the left-hand side.
16
If Windows
is used with non-Latin character sets (e.g.
®
Chinese), the area settings of the system control must be
changed to English (USA), since otherwise, communication
problems might occur.
WIKA operating instructions transmitter, models GDT-20, GDHT-20
Valid values
16 characters in ASCII code
1 ... 247
1,200 ... 115,200
None, even
0 ... 100