Herunterladen Inhalt Inhalt Diese Seite drucken

Menu Other Options - Gicam WIN11 Installations- Und Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

he analogue output is obtained with the conversion of the digital value of the gross or
net weight (DAC) with a resolution of 16 bits. (65535 divisions).
Offset adjustment (calibration)
oFFSEt
Measure the analogue output value with a tester to perform the calibration of zero (0)
and full scale (FS). Use the arrow keys to adjust the analog output. Keep the button
pressed for a quick change.
Test analog output
AnTEST
With this procedure it is possible to verify the operation of the analog output, determin-
ing the output value through the use of the keyboard. The figure on the right indicates
the percentage of the output value with respect to the full scale.

Menu other options

 
+
3 sec.
opTion
CoNSt
lAnGUA
LoCK
ACCCod
Display and printer message language
LANGUA
Select language from those available (Italian, English).
Keyboard lock selection
LOCK
It is possible to block the use of the keyboard at multiple levels:
Set-up menu access code
ACCCod
To protect access to the instrument set-up menu it is possible to set a 4-digit numeric
password. If the value 0 is set, the password is deactivated. It is advisable to program
a password if you intend to activate a keypad lock mode.
Set password
CAlibr
PArAM
EnGl
itAl
FrEE
ModSEt
Set value
FREE: No limitation.
ModSET: Limit access to the threshold set-up menu
SET: Limit to threshold set-up and threshold programming.
FUnZoP: Total lock of the keyboard.
Pagina –
page
In out
SEriAl
AnAloG
SEt
Fun2oP
Seite
75
Exit menu

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für Gicam WIN11

Inhaltsverzeichnis