Dräger Medical RespiCare VV Gebrauchsanweisung Seite 85

Heimbeatmungsgerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Trigger level
Min. volume
Paw low alarm
Synchronized intermittent mandatory
ventilation mode SIMV:
Frequency
Tidal volume
Backup rate
Flow
High pressure
Inspiratory pressure support
Trigger level
Apnea time delay
Backup tidal volume
Alarms
Low pressure alarm
High pressure alarm
Mains power failure alarm
Battery empty alarm
Frequency alarm
High pressure alarm
Inflation volume alarm
Slip alarm
Safety stop alarm
EPROM inop.
–3 to +8 mbar
Adjustable in 1 mbar increments
0.1 to 1.6 L
12 to 50 mbar
Adjustable in 1 mbar increments
4 to 20 1/min
Adjustable in 1/min increments
0.1 to 1.6 L
Adjustable in 0.01 L increments
6 to 22 1/min
Adjustable in 1/min increments
0.4 to 1.5 L/s
Adjustable in 0.1 L/s increments
14 to 60 mbar
Adjustable in 1 mbar increments
OFF, 5 to 40 mbar
Adjustable in 1 mbar increments
–2 to +5 mbar
Adjustable in 1 mbar increments
5 to 40 s
Adjustable in 1s increments
0.05 to 1.5 L
Adjustable in 0.05 L increments
2 to 35 mbar
Adjustable in 1 mbar increments
12 to 60 mbar
Adjustable in 1 mbar increments
Alarm in the event of mains power failure or
DC supply failure
Alarm if the "battery empty" symbol appears in the display
NO, 5 %, 10 %, 15 %, 20 %
Alarm if the current frequency is 5 % (10 %, 15 %, 20 %)
higher than the set frequency
Alarm if the maximum safe pressure of 75 mbar is
reached
Alarm if there is no air intake
Alarm if the bellows motor fails
Alarm in the event of device failure
Alarm in the event of an internal device error
Technical Data
85