EN-18
4.5.4
Measuring the Density of a liquid using a Calibrated Sinker (not supplied)
To enable this feature, enter the Density Setup menu and select the following; Density Type: Liquid.
Note: when the Density Type is set to Liquid, the Liquid type and Porous material selections are disabled.
Confirm that the default value displayed (sinker volume) is
correct. To edit the default values, touch Sinker Volume.
To adjust the Sinker volume value, touch the Sinker Volume
button.
A numeric input window is displayed.
Key in the desired value, then press Save.
The display returns to the previous screen with the new value
highlighted.
To return to the Density Determination home screen,
touch Back.
Press Start to start the process.
Step 1 of 2 – Weigh the Sinker in Air.
Follow the screen instructions, then press Accept
to store the sinker weight ("in air").
The DENSITY DETERMINATION – LIQUID Home screen
Main Display Line
Reference Fields
Function Buttons
ADVENTURER BALANCES
Application Icon