Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Safety

Work safety

Always use the prescribed personal protective equipment (PPE) such as protective
clothing, safety goggles, protective gloves, safety shoes, etc.
Do not operate any other devices in the immediate vicinity of the device ...
which can generate electromagnetic fields in the frequency range between
9 × 10
Hz to 3 × 10
3
which generate emission or radiation sources in the frequency range
3 × 10
Hz to 3 × 10
11
1,000 μm to 0,1 μm),
which generate ultrasonic or ionizing waves.
Do not operate the unit when adiabatic compression or shock waves may occur
(pressure wave ignition).
Do not use substances that could release energy in an uncontrolled way and cause
a pressure increase (exothermic reaction, spontaneous ignition of dusts).
Do not process hard, brittle materials such as stones, soil samples, etc., that could
destroy the evaporation flask.
Only use heating bath media that guarantee sufficient heat transfer.
Do not operate the device with overpressure.
Do not expose the glass components to a pressure difference of more than 2 bar.
Ensure that the coolant overpressure does not exceed a level of 2 bar.
Ensure that the flow velocity does not exceed 1 m/s when aspirating liquids with
flammable components (electrostatic charge, danger of ignition!).
Avoid the formation of explosion group IIC gases and potentially explosive distil-
lation residues.
Rectify malfunctions or faults on the device immediately.
Shut down the device and disconnect it from the power supply if it
is not possible to eliminate the disturbance or rectify the fault.
In such a case, contact the responsible technical service
department.
Hz,
11
Hz (in the optical spectral range wavelengths from
15
88

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis