Test point for measuring water pressure
The left hand holder on the front injector manifold can be used as the
test point for measuring water pressure for performance checks and
for validation purposes in accordance with EN ISO 15883.
To measure water pressure replace the existing fixture with a Luer
Lock adapter, e.g. E 447.
en - Areas of application