Step-by-Step Instructions
6.4
Getting Started
(1)
(2)
(3)
(4)
(5)
The cursor on the tooth will stop blinking (accompanied by a double
beep signal).
B EN PIIL DFU WEB / Rev.07 / 02-2018 (Old ZF 190292.EN)
Disconnect the charger from the device if connected.
Before connecting the measurement cable with attached lip
clip and connection hook to the patient, plug measurement
cable into the device receptacle and switch the device On –
after a short logo presentation, the main screen is displayed
and the cursor on the tooth image will start blinking.
(While turning the device On without the measurement cable
– the main screen will be displayed without cursor on the tooth
image).
Attach the lip clip to the patient.
Gently insert the file into the canal – (to ensure precise
measurements the file size should be adjusted to the canal
diameter).
Connect the hook to the file.
13/48