Setting the size of the measured area (pre-selection)
8
• In SETUP mode, select the option for presetting the measurement area. The "AREA" symbol
will be displayed.
• Press the
preset area for the current memory slot.
• Press the
memory slots for different areas: S-1 S-2 S-3 S-4 S-5 and S-6.
• Press the
to S-6.
• Press the
flash.
• Press the
• Press the HOLD button
the
• Press the button (15) to confirm and save the settings. The current setting is indicated on
the LCD display (7).

10. Measuring the sound level

Exercise caution in loud environments. You may suffer hearing damage. Always use hearing protection in
loud environments! Ensure that there are no objects or people between the measurement microphone (9)
and the sound source.
Sound sources must always be measured directly. Point the sound level meter and microphone directly at
the sound source.
To avoid influencing the sound waves, keep the meter as far away from your body as possible using an outstretched
arm, or attach it to the enclosed tripod. If possible, move away from the measuring point. There is a tripod socket (10)
on the back of the meter.
Avoid vibrating or moving the meter.
When taking measurements in windy environments (>10m/s), use the enclosed wind deflector to prevent wind noises
from resulting in incorrect measurements. The wind deflector does not distort the measurement results.
Follow the steps below to measure the sound level:
1
1. If necessary, place the wind deflector on the microphone.
2. Press the on/off button
3. Press the
button (3) to select sound level mode. When you reach sound level mode, the "SOUND" symbol will
appear on the LCD screen (7) near the display (B).
38
(15) button to display the area and area unit in display B. Display A shows the
(13) or
(14) button to select a memory slot for the area. You can configure 6
(13) or
(14) button to set the sizes of the measured areas in a loop of S-1
button (15) to select a digit from the current area setting. The selected digit will
(13) or
(14) button to set the digits from 0-9.
(4) to go to the next digit (up to four places from right to left). Press
(13) or
(14) button to set the next digit.
(1) to switch on the meter.