6. Changing the batteries
When the LCD screen displays
and replace the used batteries taking care to observe polarity. A polarity error may
cause damage and void the guarantee. Do not use batteries other than the batteries
mentioned, do not recharge non-rechargeable batteries, or throw them into a fire.
Remove the batteries when not using the device for a long period. Use disposable batteries.
VIII. PRECISION
ACCORDING
TO ASTM
STANDARD
1965-1998
(2009)
The THERMOFLASH® LX-260TE can take temperature measurements below 32°C / 89.6°F or
above 42.9°C / 109.22°F. Outside these temperature ranges, precision is not guaranteed.
Longevity of use
The THERMOFLASH LX-260T EVOLUTION is guaranteed for about 40,000 readings.
IX. ACCESSORIES SUPPLIED
User manual, Carrying bag, Batteries supplied LR6 (AA).
X. TROUBLESHOOTING
If you experience any of these problems using your THERMOFLASH® LX-260T EVOLUTION,
please refer to this troubleshooting guide to resolve the problem. If the problem persists, please
contact our customer service at +33 892 350 334 (0.40€ per minute).
- The display indicates temperatures above 95°: the temperature is indicated in Fahrenheit.
Change the unit of measure to Celsius.
- The screen displays temperatures below 32°C (89.6°F) : to measure body temperature, the
mode BODY must be selected. If the thermometer is in SURFACE or ROOM mode, the displayed
reading indicates the external temperature released by your body.
- The screen displays the message HI : the analysed temperature is above the range of the
selected mode, i.e. greater than 42.9°C (109.2°F) in BODY mode, or 60°C (140°F) in SURFACE
mode.
- The screen displays the message LO : The analysed temperature is below the measuring range
of the selected mode, i.e. less than 32°C (89.6°F) in BODY mode and less than 0°C (32°F) in
SURFACE mode.
, the battery is used.Operation: Open the cover
34°C - 35.9°C =
± 0.3°C
36°C - 39°C = ± 0.2°C
39°C - 42.5°C = ±
0.3°C
IFU_G01079_ML_ V02202006R02
93.2°F - 96.6°F = ± 0.3°F
96.8°F - 102.2°F = ± 0.2°F
102.2°F - 108.5°F = ± 0.3°F