Herunterladen Inhalt Inhalt Diese Seite drucken

Installation; Reprocessing Transmission Instruments - Miele A 315 Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 17
en - Areas of application

Installation

Reprocessing
transmission
instruments
42
An A 801 adapter is required.
Unscrew the upper part  from the AUF 1 holder.
Fit a silicone adapter (ADS)  in its place which has the same dia-
meter as the transmission instrument being reprocessed.
 Fitting a transmission instrument into a silicone adapter that is
too big can cause damage to the filter plate.
Remove an injector nozzle from the injector manifold and replace it
with the A 801 adapter .
Screw the lower part of the holder  into the adapter.
Insert a filter plate  into the lower part.
 Do not insert a filter plate into the holder if the injector manifold
has been fitted with an A 800 tubular filter.
Screw the upper part into the lower part.
Before loading check that the silicone adapters are present and
correctly fitted.
Wipe away any deposits from the area around the silicone adapters
that may be present depending on local water conditions using a
damp cloth.
Check that internal sections of transmission instruments are free of
blockages.
Fit the instruments into their holders using the respective silicone
adapters.
After reprocessing, the insides of the instruments must be dried
with sterile compressed air before being cleaned and sterilised as
appropriate in accordance with the manufacturer's instructions.
Please observe national regulations.
Before using transmission instruments again following repro-
cessing, a function check must be carried out, e.g. by spraying into
a basin, to ensure they are clear.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis