en - Areas of application

Installation

40
 The wash water used for reprocessing narrow-lumened instru-
ments needs to be filtered. The injector rails can each be fitted with
an A 800 tubular filter for this purpose. The tubular filter is supplied
with its own operating instructions.
Due to their length, surgical suction devices used in dental applica-
tions cannot be reprocessed using the A 315 module if an upper
basket is being used in the cleaning machine at the same time. For
this reason, surgical suction devices must be reprocessed in an up-
per basket with an injector rail, such as the A 105/1.
Cleaning the A 800 tubular filter
The tubular filter must be cleaned at least once a week.
We recommend cleaning the tubular filter after every 10 programme
sequences.
Follow the cleaning instructions in the operating instructions for the
tubular filter.
A 803 holder for transmission instruments
A 803 holders must be retro-fitted for reprocessing dental transmis-
sion instruments. The holders are suitable for almost all standard
transmission instruments.
Remove a nozzle from the injector manifold and replace it with an
A 803 holder.
When the A 803 holders are used, the wash water must be filtered.
The injector manifold must be fitted with an A 800 tubular filter.
The tubular filter is supplied with its own operating instructions.
Follow in particular the instructions on how to perform regular
cleaning on the tubular filter.