Error messages and malfunctions
Error
Possible causes
E-01
Adjustment error:
Measuring cell not filled with
water
Air bubbles in measuring cell
Sample tube defective
Measuring cell is dirty
Measuring cell defective
E-02
Measuring cell fault
Measuring cell is dirty
Measuring cell defective
E-03
Sample temperature not
measured correctly
E-04
Error in ambient temperature
measurement
E-05
Full appears instead of a
sample number:
Data memory full
E-06
Memory fault
E-07
Measuring time of 10 minutes
exceeded
BATT
Batteries empty
Result = 0.0000 or strongly
negative results:
Menu setting incorrect
10
Cleaning and maintenance
10.1
Cleaning the measuring cell
The built-in sample pump or an external syringe can be used to fill the measuring
cell with cleaning liquid.
- Clean measuring cell daily with a suitable cleaning liquid.
- For a very dirty measuring cell let the inside of the cell soak with cleaning
liquid.
- Repeat cleaning if necessary.
10.2
Cleaning the housing
- Never use aggerssive liquids or solvents to clean the housing of the Densito 30P !
- We recomend that you use the cleaning tissues which come with the instrument.
METTLER TOLEDO Densito 30P
manuals search engine
Error messages / Cleaning
What to do
- Fill measuring cell with water
- Use bubble-free, distilled water
- Replace sample tube
- Clean measuring cell
- Call METTLER TOLEDO Service
- Clean measuring cell
- Call METTLER TOLEDO Service
- Call METTLER TOLEDO Service
- Call METTLER TOLEDO Service
- Delete data from memory
- Call METTLER TOLEDO Service
- Switch instrument off and
on again
- Measure using bubble-free,
distilled water
If error appears again:
- Call METTLER TOLEDO Service
- Replace batteries
(see Chapter 4.1)
- Set Calib. Mode to off
21