Transferring a result
– While pressing the MEMORY OUT key, call up, e.g. No. 08 using the
– Press the MEMORY OUT key again: Result No. 08 is transferred. The display shows the next
higher sample or result number.
Transferring all results
– Press the ENTER and MEMORY OUT keys simultaneously: All results are transferred; the
result number is automatically incremented.
7.
Error messages
Messages appear in the display for several faults and malfunctions; causes and corrective
measures are shown for these and for a possible wrong temperature display.
Error message
err 1
err 2
err 3
err 4
err 5
BATT
DA-110M
METTLER TOLEDO
Cause
Error in the calibration:
1. Measuring tube not filled with H
2. Bubbles in measuring tube
3. Connection tube defective
4. Substance in measuring tube interferes
5. Sensitivity of measuring tube adversely
affected
6. Measuring tube broken
Connection to measurement unit
faulty
No oscillation of measuring tube
Sample temperature not measured
(temperature sensor faulty)
Ambient temperature not measured
(temperature sensor faulty)
Data memory full: FL appears in
place of a sample number
Batteries discharged
Measure
O
1. Add H
2
2. Use dist. H
3. Replace connection tube
4. Clean measuring tube
5. Replace measurement unit
6. Replace measurement unit
Check connection cable
Replace measurement unit
Call
METTLER TOLEDO
Call
METTLER TOLEDO
Clear memory
Replace batteries (see Section
8.1)
or
key.
O to measuring tube
2
O free from bubbles
2
service
service