Fault, Cause, Remedy
Error message in
Cause
the display
» Calibration error «
Calibration not
successful.
» Accuracy check
Calibration test not
failed «
successful.
» Sensor failure «
Sensor not within
the specification.
» Error datalogger «
Data storage not
possible.
» RFI error «
Electromagnetic
fault is too strong.
» GPS – position not
No free visibility to
found «
the satellites.
44
Remedy
Error message in
the display
Calibrate device
» Internal System
again; by trained
error «
service personnel,
with fault number
e.g. from Dräger
and text
Safety.
Calibrate device; by
trained service
personnel, e.g. from
Examples:
Dräger Safety.
Switch the device off
and on again. If the
error message
. . . heating system
appears again: have
the sensor replaced
by trained service
. . . EC-sensor (offset) Device is not ready
personnel, e.g. from
Dräger Safety.
. . . sampling system
Read the data
memory and
re-initialise; by
trained service
personnel, e.g. from
Dräger Safety.
. . . backup battery
Correct fault source
and repeat
. . . GPS
measurement.
communication
Move to a location
with free visibility and
hold the device with
the display directed
upwards.
Cause
Remedy
Device fault.
Remove batteries for
a short period of time
(2 seconds) and
reinsert (only
possible with alkaline
batteries or NiMH
rechargeable
batteries). If the error
continues to
occur:make a claim
with Dräger Safety
Service.
Device is not ready
Call Dräger Safety
to measure
Service.
Call Dräger Safety
to measure
Service.
Frequency
Call Dräger Safety
determination not
Service.
successful.
Pressure values not
OK.
RTC Error
Call Dräger Safety
Service.
Module Error
Call Dräger Safety
Service.
Dräger Alcotest 7510