DISINFECTION, CLEANING AND STERILISATION
A2. Implantable devices:
Dentinal and radicular posts made of steel, titanium and glass fibers.
Supports, kits and organiser systems for posts.
A3. Contra angle
B. Filling material: Only chemical disinfection (no sterilisation)
Gutta percha, Thermafil obturation devices.
Exclusion
Equipment such as Motors, Apex locators and other devices with
reprocessing procedures included in the individual Direction for Use.
MTA, Glyde, TopSeal.
General recommendation
1. Use only a disinfecting solution which is approved for its efficacy
(VAH/DGHM-listing, CE marking, FDA approval) and in accordance
with the DFU of the disinfecting solution manufacturer. For all metal
instruments, it is recommended to use anticorrosion disinfecting and
cleaning agents
2. For your own safety, please wear personal protective equipment
(gloves, glasses, mask).
3. The user is responsible for the sterility of the product for the first cycle
and each further usage as well as for the usage of damaged or dirty
instruments where applicable after sterility.
26/40