STEP-BY-STEP INSTRUCTIONS
Note
The temperature setting on the console has been optimized based on
clinical review. Therefore, the setting does not directly correlate to the
absolute temperature of the heated plugger.
Caution
Do not stay in the root canal with a heated tip for more than 4 seconds for
safety!
Caution
Place gutta-percha on the tip of the thermal response tip before applying
the tip to the patient's tooth! Do not place the tip directly on the tooth!
Caution
Always press the thermal response key to set the temperature to 90°C
before starting the pulp sensitivity test!
6.3.8

Operation - Flow

Dispense gutta-percha: Press the activation cuff to start the flow of filling
material. You will notice a short delay as the plunger engages and pushes
gutta-percha to the tip of the needle. Extrude a small amount of gutta-
percha from the needle. Wipe the excess gutta-percha from the tip before
inserting the needle into the canal. Engage the needle within the root
canal in accordance with your preferred technique. Hold the hand piece
lightly when placing material to allow the device to readily back out of the
canal. As the material is expressed, the indicator will help you estimate
how much material remains in the cartridge.
Note
Applying excessive downward pressure or not allowing the cannula to
back-out of the canal may result in a broken cannula.
6.3.9

Standby Mode Flow

After 20 minutes of inactivity, the Flow heater will turn off and slowly cool
to room temperature. The Temperature LED will also go out. Press any
button on the Flow side of the console to reactivate the Flow heater.
BENCALDDFUWEB / Rev.04 / 02-2019 (Old ZF 1902103.EN)
15/40