STEP-BY-STEP INSTRUCTIONS
6.4.1

Pack

Pluggers and Thermal Response Tip
Keep the connector portion of the plugger clean. Oxidation from repeated
autoclaving should be gently removed. This maintenance will insure
proper electrical connection to the hand piece.
Pluggers and tips must be cleaned, disinfected and sterilized before every
use:
1. Gently use a brush for removal of debris. Clean the disinfected
pluggers and tips under running water for a minimum of 1 minute.
Clean with water and mild, non-abrasive, detergent or disinfection
solution. Do not use disinfecting solutions which contain phenol,
anticorrosive, very acidic or very alkaline solutions. Clean the
disinfected pluggers and tips under running water for a minimum of
1 minute to remove all chemicals. Dry thoroughly.
2. Steam autoclave plugger tips for 18 minutes at 134°C before initial use
and between each patient use.
3. Observe that the pluggers and thermal response tip you are using are
not damaged. A deformed or oxidized tip should be replaced. All tips
will gradually lose their heating efficiency over a period of time.
4. The tips can be re-used, if they are treated with care, are not damaged
or contaminated. Any further use of damaged or contaminated tips is
at the user's own risk. All risk and liability is excluded in this case.
Note
Observe the concentration of the cleaning or disinfecting solution stated
by the manufacturer!
Warning
Do not touch hot pluggers!
BENCALDDFUWEB / Rev.04 / 02-2019 (Old ZF 1902103.EN)
19/40