generating equipment. To confirm ground integrity, use a commercial branch
circuit analyzer. If adverse power conditions exist, a dedicated power circuit or
power line conditioner might be required.
FOR CONTINUED PROTECTION AGAINST SHOCK HAZARD CONNECT TO
PROPERLY GROUNDED OUTLET ONLY. DO NOT REMOVE THE GROUND
PRONG.
Power Requirements
The terminal requires 100 to 240 VAC (at 400 mA maximum) with a line
frequency of 49 to 61 Hz of power.
Ethernet and USB Connections
The IND780's Ethernet connection provides a 10/100 base T connection (10/100
Mb) via a standard RJ45 connector.
The standard ST30 USB connector allows USB-supported peripherals, such as
keyboards, to be attached to the IND780.
Figure 2-30 shows the locations of the Ethernet and USB connectors on the main
PCB. In this case, the PCB is mounted in a Panel enclosure.
Figure 2-30: Ethernet and USB Connections
COM1 and COM2 Serial Port Connections
The COM1 port includes connections for RS232, while COM2 supports connections
for RS-232, RS-422 and RS-485. The Interface parameter (in Setup at
Communication > Serial) must be set to match the hardware connection used.
This parameter controls how the transmit and receive lines are controlled.
COM1 Serial Port
Figure 2-31 indicates which terminal represents which signal on the COM1 Port,
and Figure 2-32 shows how the port is wired for an RS232 connection. Make the
IND780 Installation Manual
WARNING
USB Connector
RJ45 (Ethernet)
Connector
2-19