Maintenance
5.1

Maintenance intervals

i
Routine maintenance ensures constant readiness for
operation.
The following maintenance intervals are to be observed
unless otherwise specified by the approval authorities.
5.1.1
Half-yearly maintenance
Replacement of coarse filter GF1 (Fig. 4, Pos. 8) in
sampling hose.
Replacement of coarse filter GF2 (Fig. 4, Pos. 1),
refer to Section 5.2.4.
Check whether both PVC hoses are connected to the
gas outlets (Fig. 4, Pos. 6 and 7).
Visual inspection of sampling probe (Fig. 4, Pos. 10).
Sampling probe leak test (refer to Section 5.2.1).
5.1.2
Annual maintenance
This work is to be performed by a qualified mainte-
nance service. It includes the half-yearly maintenance
and the following additional items:
Checking of BEA 060 measurement accuracy with
a test gas.
Replacement of activated charcoal filter
(Fig. 4, Pos. 4) in zero gas path.
Replacement of coarse filter GF3
(refer to Section 5.2.4)
Compliance with the specifications of the relevant
authorities must be ensured.
Robert Bosch GmbH
N0
11-16V
20 W
459907_4Ko.eps
7
11
Fig. 3:
BEA 060
1 Coarse filter GF2
2 Coarse filter GF3
3 8 m hose
4 Activated charcoal filter AF1
5 O
sensor
2
6 Measurement gas outlet (transparent PVC hose)
7 Gas outlet and condensate outlet (transparent PVC hose)
8 Coarse filter GF1
9 30 cm Viton hose (black)
10 Exhaust gas sampling probe
11 Plastic hose for leak test
5.2
Maintenance
5.2.1
Freedom from leaks of sampling system
To ensure accurate exhaust gas measurements it is es-
sential that there are no leaks in the sampling system
(exhaust gas sampling probe, hoses, coarse filters). It is
therefore advisable to conduct the leak test every day.
5.2.2
Exhaust gas sampling probe
Keep the opening at the tip of the exhaust gas sampling
probe clean. In the event of HC residue and conden-
sate, detach the exhaust gas sampling probe from the
hose and blow compressed air through it in the direc-
tion opposite to intake direction.
5.2.3
Sampling hose
Check for damage. In the event of HC residue or con-
densate, detach the hose from the analyzer and blow
compressed air through it in the direction opposite to
intake direction.
| BEA 060 | 27
en
BEA 060
GF 3
GF 2
O2
AF I
1
2
3
6
5
4
10
|
1 689 989 112
2016-05-12
8
9