Herunterladen Inhalt Inhalt Diese Seite drucken

Data Transmission - Seca 655 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für 655:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Fault/error code
083-297XX-XXX
084-297XX-XXX
099-297XX-XXX

10.4 Data transmission

Fault/error code
The "Date of birth" dialog
window appears after
pressing the
key
Autoclear function: Entering
"0 seconds" is not accepted;
instead, the factory setting
(300 seconds) is suggested
Workflow LED does not
light up
128 •
seca 655/654
Cause
An error occurred during the calibration
procedure.
Interference caused by reflection
Interference caused by other ultrasonic
emitters
The ambient temperature is too high or
too low.
The temperature sensor is defective.
Voice guidance: Language that does
not support the announcement of
measuring results in imperial units
active.
Announcement of the measuring
results was automatically deactivated.
Cause
Patient ID not scanned
ID scanning not set up
Patient does not yet have a file in the
EMR System
Device is connected to the
seca analytics 125
software
Implausible input; switching off the
function is not intended in the factory
No network connection
WiFi function deactivated
Workflow LED is defective
Remedy
• Remove objects from the immediate vicin-
ity of the device ➔
Setting up the device,
page
93.
• Ask persons in the vicinity to remain
further away from the device ➔
the device, page 93
Increase the distance from other ultrasonic
emitters ➔
Setting up the device, page 93
• Observe ambient conditions for operation,
transport, and storage ➔
technical data, page 135
Inform seca Service
Set metric units and activate
announcement of the measuring results
Switching over units, page 114
Activating/deactivating the
announcement of measuring results
(Results), page 121
Select language that supports
announcement of the measuring results in
imperial units (EN (US/UK), ES/ES-MX) and
activate announcement of the measuring
results.
Selecting a language, page 119
Activating/deactivating the
announcement of measuring results
(Results), page 121
Remedy
• Press
key
• Scan patient ID
Check workflow settings for the device in
the
software
seca connect 103
• Enter date of birth and press
again
• Create patient file in the EMR System and
assign measurement
• Enter date of birth
• Press
key
• Accept factory setting
• Manually enter value between 1 and 3600
seconds
Set up network connection ➔
network functions, page 115
Activate WiFi function ➔
Activating/
deactivating the WiFi function, page 117
Inform seca Service
Setting up
General
key
Setting up

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

654Mbca 555Mbca 554Mbca 552

Inhaltsverzeichnis