Power frequency
(50/60 Hz)
IEC 61000-4-8
Note: UT is the A.C. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration — electromagnetic immunity
The VRA201 is intended for use in the electromagnetic environment specified below. The customer or the
user of the VRA201 should assure that it is used in such an environment,
Immunity test
Conducted RF
IEC / EN 61000-4-6
Radiated RF
IEC / EN 61000-4-3
VRA201 - Instructions for Use - EN
30 A/m
IEC / EN 60601 test
level
3 Vrms
150kHz to 80 MHz
6 Vrms in ISM bands
150kHz to 80 MHz
80 % AM at 1 kHz
3 V/m
80 MHz to 2,7 GHz
80 % AM at 1 kHz
30 A/m
Compliance level
3 Vrms
6 Vrms
3 V/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or residential
environment.
Electromagnetic environment –
guidance
Portable and mobile RF
communications equipment
should be used no closer to any
parts of the VRA201, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance:
d
1
2 ,
P
=
2 , 1
d
=
P
80 MHz to 800 MHz
d
2
3 ,
P
=
800 MHz to 2,7
GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
should be less than the
compliance level in each
frequency range.
b
Interference may occur in the
vicinity of equipment marked with
the following symbol:
a
Page 16