5) Insert the plug connection with connected electrode cable as far
as possible into the product (see fig. 3).
6) Remove any silicone grease that seeps out.
Electrode cable replacement
1) Pull off the electrode cable at right angles to the product (see
fig. 4).
2) Cut off at least 5 mm of the electrode cable and push it back
into the plug connection.
6.2 Inserting the product
► Insert the product into the inner socket from the outside.
For slipping the electrode into the mounting channel, slightly
press the suspension arms together. The suspension arms
engage if slight pressure is applied.

6.3 Adjusting the product

1) Use the adjustment tool to set the product sensitivity (see fig. 5).
2) After a pause of 10 to 15 minutes, readjust the sensitivity.
INFORMATION
The patient must be able to hold the maximal value of the control
unit for at least 2 seconds. If this is not observed, the product set­
ting is overly sensitive which makes it more difficult for the patient
to control the prosthesis.
INFORMATION
The scale values of the 13E200 electrode do not match the scale
values of the 13E125 electrode. Therefore, do not adjust the elec­
trodes by comparing positions of the two electrodes.