4.6
By default the VSM72 performs a continuous transition between Pirani and cold
cathode range whereupon an assimilation of the sensor signals is carried out.
In order to adapt the performance of the transducer to the requirements of the
vacuum process you can configure the parameter "mode sensor transition"
according to the following:
0:
1:
Address:
Code:
Data:
cks:
CR:
Example (for address-switch in position "1")
Action
Set mode sensor transition
to continuous transition
Read mode sensor transition
Please find further information about the Thyracont communication protocol in
chapter 4.2.
Downloaded from
www.Manualslib.com
Mode Sensor Transition
no transition, but direct switch-over between Pirani and cold cathode at
-3
1.0 x10
mbar
continuous transition in the range 1.0 – 2.0x10
Address
Code
3 Bytes, decimal; address space RS485: 001 – 999
"W" for reading, "w" for writing the parameter
6 Bytes; "000001" transition, "000000" direct switch-over
1 Byte, checksum, defined as sum over all Bytes of the fields
address, code and data, modulo 64 plus 64.
Carriage Return (0Dh, 13d)
manuals search engine
Data
cks
:
Telegram to
transducer
C
"001w1k
"
R
C
"001Wh
"
R
-3
mbar (standard)
CR
Answer telegram
from transducer
C
"001w1k
"
R
C
"001W1Y
"
R
(
transition)
39