9. Specifications
9. Specifications
Model
Rated force F
kN
nom
Relative linearity error d
Relative reversibility error
Relative creep, 30 min. at F
Temperature effect on
characteristic value TK
zero signal TK
0
Limit force F
L
Breaking force F
B
Permissible vibration loading F
Rated displacement (typ.) s
<10 kN
<100 kN
Material of the measuring body
Rated temperature range B
Service temperature range B
Storage temperature range B
Electrical connection
Output signal
(Rated characteristic value) C
WIKA operating instructions, models F2301, F23C1, F23S1, F2303, F2304
F2301
1, 2, 3, 5, 10, 20, 30, 50, 100,
200, 300, 500
1)
±0.5 % F
lin
< 0.1 % F
0.1 % F
nom
0.4 % F
c
0.4 % F
150 % F
300 % F
±50 % F
rb
nom
< 0.02 mm
< 0.2 mm
Corrosion-resistant stainless steel, ultrasound-tested
3.1 material (optionally 3.2)
-20 ... +80 °C
T, nom
-30 ... +80 °C
T, G
-40 ... +85 °C
T, S
Circular connector M12 x 1,
4-pin / CANopen
4 ... 20 mA, 2-wire
4 ... 20 mA, 3-wire or
nom
DC 0 ... 10 V, 3-wire
(Optional redundant signal)
CANopen
accordance with CiA 301,
device profile 404, communi-
cation services LSS (CiA 305),
configuration of the instrument
address and baud rate Sync/
Async, Node/Lifeguarding,
heartbeat; zero and span
±10 % adjustable via entries in
the object directory
nom
nom
nom
/10 K
nom
/10 K
nom
nom
nom
(in accordance with DIN 50100)
nom
®
5-pin
®
protocol in
2)
F23S1
3, 5, 10, 20, 30, 50, 100
2-connector version
M12 x 1, 4-pin
Redundant, opposing,
4 ... 20 mA /
20 ... 4 mA, version in
acc. with requirements
for functional safety per
Machinery Directive
2006/42/EC
27
EN