OBJ_DOKU-184279-008.fm Seite 25 Montag, 5. November 2018 8:20 08
5

Unit settings

WARNING! Sparks! If the voltage or power is too high, sparks
may be generated around the jaws of the instrument. This may
cause irreparable tissue damage and may damage the product.
• Select the device settings as specified in the notes on use.
Do not use an AUTO-START function.
5.1

Erbe ICC 350

• Program: "b"
• Recommended CUT power: 65 watts
• Recommended CUT effect: 2 (Do not exceed effect 3!)
• Recommended COAG power: 50 watts
5.2

Erbe VIO S and Erbe VIO D

CUT settings:
• BIPOLAR CUT mode
• Power limiting: 80 watts
• Recommended effect: 3 (Do not exceed effect 4!)
COAG settings:
• BIPOLAR SOFT COAG mode
• Power limiting: 60 watts
• Recommended effect: 4