Herunterladen Inhalt Inhalt Diese Seite drucken

Maintenance And Care - Integra MAYFIELD A1015 Gebrauchsanleitung

Crossbar adaptor
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 29
EN – ENGLISH
Phase
Time
Water Tem-
(Min.)
perature
Wash 1
10:00
60.0°C
(140° F)
Rinse 1
0:30
Hot water
Thermal
2:00
82.2°C (180°
Rinse**
F)
** Optional phase for disinfection of components –
minimum water temperature as indicated or per worker
manufacturer specifications for the thermal rinse cycle.
NOTE – Any deviation from this guideline could result in
damage to the equipment as well as improper cleaning
results.
Rinse with purified water. Do not perform if parameters
cannot be achieved.
5. Remove from washer and dry completely if needed.
6. Inspect equipment to make sure there is no visible
organic debris or residue from cleaning agent.
Repeat process if any visible soil is detected.

Maintenance and Care

To ensure proper function and to extend the life and
performance of the equipment, Integra LifeSciences
recommends the following:
Recommended Action
Return the device to the Integra
LifeSciences Repairs department for
detailed inspection and servicing.
Request that Integra NeuroSpecialists
perform routine inspections of the
device
In the absence of proper care and servicing of the device,
negative effects may be seen after repeated processing
over time which may lead to reduced performance.
Contact information: See the Service and Repair section
for contact information on how to return your device for
periodic servicing and to request periodic inspections.
Detergent and
Concentration
Neutral pH
detergent
(e.g. Renu-Klenz™
, Steris, Ratio
1:256)
N/A
N/A
Recommended
Frequency
Once / year
Twice / year
8
See Inspection and/or Service notes section for routine
checks to be performed on the device.
Device Disposal
NOTE: Follow hospital procedures for disposal of this device.
Integra LifeSciences Warranty Statement
INTEGRA LIFESCIENCES CORPORATION ("INTEGRA")
warrants to the original purchaser only that each new
MAYFIELD product is free from manufacturing defects in
material and workmanship under normal use and service
for a period of one year (except as otherwise expressly
provided as to accessory items) from the date of delivery
by INTEGRA to the first purchaser, but in no event beyond
the expiration date stated on any product labeling.
• Surgical instruments are guaranteed to be free
from defects in material and workmanship when
maintained and cleaned properly and used normally
for their intended purpose.
• Any covered product that is placed by INTEGRA
under a lease, rental or installment purchase
agreement and that requires repair service during
the term of such placement agreement shall be
repaired in accordance with the terms of such
agreement.
If any covered defect occurs during the warranty period
or term of such placement agreement, the purchaser
should communicate directly with INTEGRA's home
office. If purchaser seeks to invoke the terms of this
warranty, the product must be returned to INTEGRA at its
home office. The defective product should be returned
promptly, properly packaged and postage prepaid. Loss
or damage in return shipment to INTEGRA shall be at
CUSTOMER's risk. INTEGRA's sole responsibility under
this warranty shall be repair or replacement, at INTEGRA's
sole discretion at INTEGRA's expense, subject to the terms
of this warranty and applicable agreements.
IN NO EVENT SHALL INTEGRA BE LIABLE FOR ANY
INCIDENTAL, INDIRECT, CONSEQUENTIAL OR PUNITIVE
DAMAGES IN CONNECTION WITH THE ACQUISITION
OR USE OF ANY INTEGRA PRODUCT. Further, this
warranty shall not apply to, and INTEGRA shall not be
responsible for, any loss arising in connection with the
purchase or use of any INTEGRA product that has been
repaired by anyone other than an authorized INTEGRA
service representative or altered in any way so as, in
INTEGRA's judgment, to affect its stability or reliability,
or which has been subject to misuse, negligence or
accident, or which has been used otherwise than in

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Mayfield a1016

Inhaltsverzeichnis